Product Description



Brand

CLEANALYZER

Application

Cleanlyzer Hematology Cleaning Kit

Contents

480 ml Blue Colored Solution Enzymatic Cleaner, 480 ml Sputo Cleaner, 480 ml Detergent Cleaner (1/Kt)


Pack Size

3 x 480 ml

Art No.

170201

Features

  • The CLEANALYZER cleaning kit is a unique triple action kit designed for use on hematology analyzers
  • The cleaning fluids will, in three steps, thoroughly clean and relubricate your analyzer, prolonging the life of the components and inimizing the risk of bacterial growth and background problems
  • Longer life time of instrument tubing
  • Lower background counts, increased mean time between failure
  • It is usual daily applied chemical cleaning solution, comprising isotonic and, in addition, surfactants, which allow to remove remnants during` washing of blood cells. It also contains the green color dye harmless, by which the solution is visible in the connecting device hoses. During normal daily operations container with a solution Detectoterge permanently connected to the device and solution is supplied as a working cleanser. Before each feeding another sample analyzer exhauster purified solution Detectoterge. Application Detectoterge solution gave excellent results with Hematology:
  • Depending on the task, Detectoterge can also be used in special purification protocols.

Cleaning Procedure :

Cleaning with CLEANALYZER- CELL CLEANING KIT p

Put both probes (D L) in the CLEANALYZER Enzymatic Cleaner (BLUE-1 No.) Perform FIL SYSTEM wait approx. 45 Min. Empty the System after put out both probes (D L) p

Put both probes (D L) in the CLEANALYZER - Sputo Cleaner (YELLOW-2 No.). Perform FIL SYSTEM wait not More than 1 Min; Empty the System after put out both probes (D L) p

Put both probes (D L) in the CLEANALYZER- Detergent Cleaner (RED-3 No.). Perform FIL SYSTEM wait approx. 20 Min. Empty the System after put out both probes (D L)

Put both probes (D L) in the Dis. Water. Perform FIL SYSTEM wait approx. 5 Min.; Empty the System after put out both probes (D L)

And finally;

Put respective probes (D L) in the Diluent Lyzer Bottles. Perform FIL SYSTEM